752 Appendix F
Replacing the 8753ES with the E5070B/E5071B
Comparing Functions
Measurement
(cont'd.)
Sweep
conditions
Number of points
(except for the list frequency
sweep operation)
3, 11, 21, 26, 51, 101, 201, 401, 801,
1601
Arbitrary value from 2 to 1601
*1
Sweep type linear sweep, log sweep, list sweep,
power sweep, and CW time sweep
linear sweep, log sweep, segment
sweep, and power sweep
List frequency
sweep
Number of list
tables
One for channels 1 and 3 and one for
channels 2 and 4, two in total
One for each channel (16 in total)
Number of
segments per
table
Up to 30 Up to 201
Creation of a
table using
GPIB(SCPI)
command
Creating a table using more than one
command
Creating a table using a single
command
Number of
points
1 to 1601 points per segment
Up to 1601 points in total
2 to 1601 points per segment
Up to 1601 points in total
*1
Stepped/swept
mode
stepped mode and swept mod.
Selectable only when list frequency
sweep is performed
stepped mode, swept mode, fast
stepped mode, and fast swept mode.
Selectable also when list frequency
sweep is not performed.
Sweep of
designated
segments.
Allowed Not allowed (Always sweeps all
segments.)
Sweep time Automatic, Manual (definable from
the shortest time to 24 hours)
Automatic, Manual (range of
definable sweep depends on sweep
condition)
IF bandwidth 10 Hz, 30 Hz, 100 Hz, 300 Hz,
1 KHz, 3 kHz, 3.7 kHz, 6 kHz
10 Hz, 15 Hz, 20 Hz, 30 Hz,
40 Hz, 50 Hz, 70 Hz, 100 Hz,
150 Hz, 200 Hz, 300 Hz,
400 Hz, 500 Hz, 700 Hz, 1 kHz,
1.5 kHz, 2 kHz, 3 kHz, 4 kHz,
5
KHZ, 7 KHZ, 10 KHZ, 15 KHZ,
20
KHZ, 30 KHZ, 40 KHZ, 50 KHZ,
70
KHZ, 100 kHz
Averaging Can be set from 1 to 999. Can be set from 1 to 999.
Smoothing The smoothing aperture can be set
from 0.05% to 20%.
The smoothing aperture can be set
from 0.05% to 25%.
Electrical delay Transmission line Compatible with coaxial cables and
waveguides
Compatible with coaxial cables only
Value Any value from 0 to 10 seconds Any value from -10 to +10 seconds
Phase offset Any value from -360° to +360°. Any value from -360° to 360°.
Setting sweep conditions for Couple/Uncouple Channels 1 and 3 and channels 2 an 4
are always coupled. Channels 1 and
2 can be set either at Couple or
Uncouple.
Traces in the same channel are
coupled; traces in different channels
are not coupled.
Trigger mode Continuous, single, hold, specified
number of sweeps
Continuous, single, hold
Table F-5 Functions of the 8753ES vs. the E5070B/E5071B
Function 8753ES E5070B/E5071B