Agilent Technologies ENA Series Guia do Utilizador Página 203

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 202
Chapter 4 197
Calibration
Changing the Calibration Kit Definition
4. Calibration
85036B/E
NOTE The designations (m) and (f) in the name (label) of the standard indicate male (m) and
female (f) analyzer connectors.
1. Short(m) 2. Open(m) 3. Broadband 4. Thru
Label Short(m) Open(m) Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0 63.5 0 0
C1 [×10
-27
F/Hz]
08400
C2 [×10
-36
F/Hz
2
]
05600
C3 [×10
-45
F/Hz
3
]
0000
L0 [×10
-12
H]
0000
L1 [×10
-24
H/Hz]
0000
L2 [×10
-33
H/Hz
2
]
0000
L3 [×10
-42
H/Hz
3
]
0000
Offset Delay [s] 0 0 0 0
Offset Z0 [Ω]75757575
Offset Loss [Ω/s] 1.13 G 1.13 G 1.13 G 1.13 G
Arb. Impedance [Ω]50507550
7. Short(f) 8. Open(f)
Label Short(f) Open(f)
STD Type Short Open
C0 [×10
-15
F]
041
C1 [×10
-27
F/Hz]
040
C2 [×10
-36
F/Hz
2
]
05
C3 [×10
-45
F/Hz
3
]
00
L0 [×10
-12
H]
00
L1 [×10
-24
H/Hz]
00
L2 [×10
-33
H/Hz
2
]
00
L3 [×10
-42
H/Hz
3
]
00
Offset Delay [s] 17.544 p 17.544 p
Offset Z0 [Ω]75 75
Offset Loss [Ω/s] 1.13 G 1.13 G
Arb. Impedance [Ω]50 50
Vista de página 202
1 2 ... 198 199 200 201 202 203 204 205 206 207 208 ... 774 775

Comentários a estes Manuais

Sem comentários