Chapter 16 559
Specifications and Supplemental Information
Measurement Throughput Summary
16. Specifications and
Supplemental Information
Measurement Throughput Summary
Table 16-20 Typical Cycle Time for Measurement Completion
*1*2
(ms)
*1.Typical performance.
*2.Sweep mode: Fast swept. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1. System error
correction: OFF.
Number of Points
51 201 401 1601
Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected 45718
2-port cal 581342
Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth
Uncorrected 11 12 13 23
2-port cal 20232546
Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected 19 24 24 24
2-port cal 37464850
Table 16-21 Typical Cycle Time for Measurement Completion
*1*2
(ms)
*1.Typical performance.
*2.Sweep mode: Fast swept. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1. System error
correction: ON.
Number of Points
51 201 401 1601
Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected 46822
2-port cal 5 10 16 56
Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth
Uncorrected 11 12 13 23
2-port cal 20242555
Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected 20 24 24 26
2-port cal 37464757