196 Chapter 4
Calibration
Changing the Calibration Kit Definition
85032B
NOTE The designations (m) and (f) in the name (label) of the standard indicate male (m) and
female (f) analyzer connectors.
1. Short(m) 2. Open(m) 3. Broadband 4. Thru
Label Short(m) Open(m) Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0119.0900
C1 [×10
-27
F/Hz]
0-36.95500
C2 [×10
-36
F/Hz
2
]
0 26.258 0 0
C3 [×10
-45
F/Hz
3
]
0 5.5136 0 0
L0 [×10
-12
H]
0000
L1 [×10
-24
H/Hz]
0000
L2 [×10
-33
H/Hz
2
]
0000
L3 [×10
-42
H/Hz
3
]
0000
Offset Delay [s] 93 f 0 0 0
Offset Z0 [Ω] 49.992 50 50 50
Offset Loss [Ω/s] 700 M 700 M 700 M 700 M
Arb. Impedance [Ω]50505050
7. Short(f) 8. Open(f)
Label Short(f) Open(f)
STD Type Short Open
C0 [×10
-15
F]
0 62.14
C1 [×10
-27
F/Hz]
0 -143.07
C2 [×10
-36
F/Hz
2
]
0 82.92
C3 [×10
-45
F/Hz
3
]
00.76
L0 [×10
-12
H]
00
L1 [×10
-24
H/Hz]
00
L2 [×10
-33
H/Hz
2
]
00
L3 [×10
-42
H/Hz
3
]
00
Offset Delay [s] 17.817 p 17.411 p
Offset Z0 [Ω] 50.209 50
Offset Loss [Ω/s] 2.1002 G 700 M
Arb. Impedance [Ω]50 50