Chapter 13 431
Setting and Using the Control and Management Functions
Setting the GPIB
13.Setting and Using the
Control and Management
Functions
Setting system controller (USB/GPIB interface) when c drive volume
label in hard disk is more than CP810
When controlling an external device from the E5070B/E5071B, connect the USB port of
the E5070B/E5071B and the GPIB port of the external device through the USB/GPIB
interface.
Follow these steps to set the USB/GPIB interface when c drive volume label in hard disk is
more than CP810:
NOTE Do not connect two or more USB/GPIB interfaces.
NOTE 82357B USB/GPIB interface can not be used if not the c drive volume label in hard disk is
more than CP810. When use the 82357B USB/GPIB interface, hard disk is requred to
change.
Step 1. Connect the USB/GPIB interface to the USB port of the E5070B/E5071B. The USB/GPIB
Interface Detected dialog box (Figure 13-4) appears.
Figure 13-4 USB/GPIB Interface Detected dialog box
Step 2. Choose the “When a new 82357 is plugged in” in Show this dialog flame, then click
OK
button.