Chapter 2 51
Overview of Functions
Screen Area: Names and Functions of Parts
2. Overview of Functions
5-10. Channel Measurement Status
Displays the update status of traces on the channel.
5-11. Error Correction Status
Displays the execution status of error correction on the channel. For details, see “Execution
status of error correction for each channel” on page 102.
5-12. Power Calibration Status
Displays the execution status of power level error correction on the channel. For details,
see “Turning ON or OFF power level error correction” on page 208.
5-13. Port Extension Status
Shows whether the port extension is turned ON or OFF.
5-14. Fixture Simulator Status
Shows whether the fixture simulator is turned ON or OFF.
5-15. Averaging Status
Displays the averaging factor and averaging count when averaging is turned on.
5-16. Sweep Range
Indicates the sweep range by using the start/stop or center/span.
5-17. IF Bandwidth/CW Frequency
Indicates the IF bandwidth when the sweep type is linear/log frequency or the CW
frequency when the sweep type is power.
! Measurement in progress. When the sweep time exceeds
1.5 seconds,
is displayed at the point on the trace.
# Invalid traces. The measurement conditions have changed,
but the traces on the channel currently displayed have not
been updated to match the new conditions.
(No display) The measurement has not been executed.
PExt (displayed in blue) Port extension: ON
(not displayed) Port extension: OFF
Sim (displayed in blue) Fixture simulator: ON
(not displayed) Fixture simulator: OFF
n/m (displayed in blue) Averaging: ON
(m: averaging factor; n: averaging count)
(not displayed) Averaging: OFF