354 Chapter 10
Data Output
Saving and Recalling Instrument State
Saving and Recalling Instrument State
You can save the instrument state of the E5070B/E5071B into a file on mass storage (hard
disk drive or floppy disk drive) and then recall it later to reproduce that state. You can
select the stored data from the following four types.
In addition, the “user-preset” function is provided to allow the user to freely set up an
instrument state recalled when the preset function is executed. For more information on
how to set it up, see “Setting the user preset function” on page 453 .
Compatibility of files related to saving and recalling
The compatibility of saving/recalling the instrument state file is as follows:
R Compatibility between different models
• Files saved with the E5070B cannot be recalled with the E5071B.
• Files saved with the E5070A can be recalled with the E5070B, although the
opposite is not possible.
• Files saved with the E5071A can be recalled with the E5071B, although the
opposite is not possible.
Table 10-1 Stored data
Type Stored data and usage
State only
(State Only)
Saves the setting
*1
of the E5070B/E5071B and reproduces the state when it was saved
by recalling it later into the E5070B/E5071B.
State and
calibration data
(State & Cal)
Saves the setting
*1
of the E5070B/E5071B and calibration data (calibration coefficient
array
*2
) to reproduce the state when it was saved by recalling it later into the
E5070B/E5071B. At this time, you can perform error correction of measured values
by using the recalled calibration data.
State and trace
(State & Trace)
Saves the setting
*1
of the E5070B/E5071B and traces (error-corrected data array
*2
and error-corrected memory array
*2
) to reproduce the state when it was saved by
recalling it later into the E5070B/E5071B. At this time, the traces are also recalled and
displayed on the screen.
State, calibration
data, and traces
(All)
Saves the setting
*1
of the E5070B/E5071B, calibration data, and traces to reproduce
the state when it was saved by recalling it later into the E5070B/E5071B. At this time,
the calibration data and traces are also recalled.
*1.For information on the saved target, refer to Appendix C, “List of Default Values.”
*2.For information on each array, refer to “Data Processing” on page 736.
Recalling
E5070B
*1
E5071B
*2
E5070A E5071A
Saved
file
E5070B
*1
Y N N N
E5071B
*2
NYN N
E5070A Y
NYN
E5071A
NYNY