596 Appendix A
Manual Changes
Manual Changes
For more information on defining a calibration kit, see ??.
3. Performing Calibration
Measure the necessary calibration data and enable error correction.
NOTE The definition of the frequency range of the line standard and match standard used for
measurement must cover the sweep range of the channel for which you perform
calibration.
If the
Thru, Reflection, Match, and Line keys do not appear, go back to the calibration kit
definition menu and confirm that the frequency range definitions of match and lines are
correctly entered to cover the ENA’s measurement frequency range.
NOTE In multiport TRL/LRM calibration, the through and line calibration measurements are not
performed for all of the paths among the test ports. The following table shows the
measured paths of the through and line calibration measurements. The through and line
calibration data of non-measured paths are mathematically derived from the through and
line calibration data of the measured paths.