Agilent Technologies ENA Series Guia do Utilizador Página 606

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 605
600 Appendix A
Manual Changes
Manual Changes
Note on use
When the fixture simulator is ON and the port impedance conversion is ON, Z0 of all ports
to be saved must be set to the same value.
Operating Procedure
1. Starting VBA macro
Step 1. Press
[Macro Setup].
Step 2. Press
Load Project.
Step 3. The Open dialog box appears. Specify the file name “D:\Agilent\SaveToTouchstone.vba”
and press the
Open button.
Step 4. Press
[Macro Run] to start the macro. (Refer to Figure A-10.)
Figure A-10 SaveToTouchstone dialog box
2. Saving data
Step 1. Select the number of ports (1 in Figure A-10) and test ports (2 in Figure A-10).
NOTE You can selects 1 port or 2 ports as the number of ports when the maximum number of
channels/traces is 16 channels/4 traces or 12 channels/6 traces.
Step 2. Select a channel (3 in Figure A-10).
NOTE The channel selected in this step has no relation to the active channel.
Step 3. Select the data saving format (4 in Figure A-10).
Step 4. Press the
Save button (5 in Figure A-10). Measurement of necessary data for the selected
channel in Step 2 starts.
NOTE Regardless of the state of the trigger system, measurement is automatically performed
real - imaginary (RI) real and imaginary parts
magnitude - angle (MA) linear magnitude and phase (degree)
dB - angle (DB) logarithmic magnitude (dB) and phase (degree)
Vista de página 605
1 2 ... 601 602 603 604 605 606 607 608 609 610 611 ... 774 775

Comentários a estes Manuais

Sem comentários