Chapter 4 189
Calibration
Changing the Calibration Kit Definition
4. Calibration
Step 7. In the same way, repeat Step 2 to Step 6 to enter the definition of REFLECT to No. 2.Select
SHORT for STD Type.
Procedure to define Match
Step 1. Press
3:No Name - Label.
Step 2. Type in
MATCH <2G.
Step 3. Press
STD Type - Load.
Step 4. Set
Max Frequency to 2GHz.
Step 5. Press
Return to return to the Define Std menu.
Procedure to define Line 1/2
Step 1. Press
4:No Name - Label.
Step 2. Type in
LINE <7G.
Step 3. Press
STD Type - Thru.
Step 4. Set
Offset Delay to 54ps.
Step 5. Set
Min Frequency to 2GHz.
Step 6. Set
Max Frequency to 7GHz.
Step 7. Press
Return to return to the Define Std menu.
Step 8. Press
5:No Name - Label.
Step 9. Type in
LINE >7G.
Step 10. Press
STD Type - Thru.
Step 11. Set
Offset Delay to 13ps.
Step 12. Set
Min Frequency to 7GHz.
Step 13. Press
Return to return to the Define Std menu.
Step 14. Press
Return to return to the Modify Cal Kit menu.
Procedure to register Thru/Reflect/Match to sub class 1
Step 1. Press
Specify CLSs - Sub Class.
Step 2. Select
Sub Class1.
Step 3. Press
TRL Thru - Set All - THRU - Return.
Step 4. Press
TRL Reflect - REFLECT.
Step 5. Press
TRL Line/Match - Set All - MATCH <2G - Return.
Procedure to register Line 1/2 to subclass 2/3
Step 1. Select
Sub Class2.
Step 2. Press
TRL Line/Match - Set All - LINE <7G - Return.