Chapter 12 407
Optimizing Measurements
Setting Port Extensions and Loss Values
12. Optimizing Measurements
Step 5. Method to set the frequencies used for calculation.
Step 6. If you have selected
User Span, use User Span Start and User Span Stop to set a start
value and a stop value.
NOTE For Current Span and User Span, a frequency point at 1/4 of the frequency range is set to
Freq1; a frequency point at 3/4 of the frequency range is set to Freq2. (See Figure 12-5.)
NOTE If the setting is not made before starting OPEN/SHORT standard measurement, it does not
affect the calculation result.
Figure 12-5 Frequency points in a frequency range (Freq1 and Freq2)
Softkey Function
Current Span Executed using the frequency range set currently.
Active Marker
Executed using the frequency at the active marker.
*1
In this case, the result
is applied to Loss1. Loss2 is ignored.
*1.Even if the active marker has been disabled, it is automatically enabled.
User Span Executed using a start value and a stop value you set.