Chapter 8 323
Frequency-Offset Measurement (Option 008)
Measurement of Mixers
8. Frequency Offset
Measurement
(Option 008)
Figure 8-8 Avoid Spurious Function
NOTE You cannot avoid all spurious signals by using the avoid spurious function.
Procedures for controlling spurious avoidance function
Follow the steps below to control the spurious avoidance function.
Step 1. Press -
Frequency Offset to display the “Frequency Offset Menu.”
Step 2. Press
Avoid Spurious to turn it ON.
NOTE Even if the frequency-offset function is turned OFF, the function is available when the
external signal source (LO Frequency) is ON.
NOTE If the Swept is selected for the Sweep Mode, it will be changed to Stepped (from Std
Swept to Std Stepped, and from Fast Swept to Fast Stepped).
4. Changing Frequency Data
The frequency-offset function offers the capability to change the frequency data of each
trace to any desired frequency.
Procedures for changing frequency data
Step 1. Press (or ) and (or ) to activate the trace for
which you want to change the frequency.
Step 2. Press -
Frequency Offset to display the “Frequency Offset Menu.”
Comentários a estes Manuais