186 Chapter 4
Calibration
Changing the Calibration Kit Definition
Step 4. Press Modify Kit. To change the pre-defined connector type (e.g. OPEN(f) to OPEN (m))
skip to Step 14.
Step 5. Press
Define STDs.
Step 6. Select the standard to be redefined from among standards numbered 1 through 21.
Step 7. Press
STD Type.
Step 8. Select the type of standard according to the list below.
Step 9. Set the standard coefficient according to the list below.
Softkey Function
Open Selects the OPEN standard
Short Selects the SHORT standard
Load Selects the LOAD standard
Delay/Thru Selects the delay/THRU standard
Arbitrary Selects the arbitrary impedance
None Selects no standard type
Softkey Function
C0 Sets C0
C1 Sets C1
C2 Sets C2
C3 Sets C3
L0 Sets L0
L1 Sets L1
L2 Sets L2
L3 Sets L3
Offset Delay Sets the offset delay
Offset Z0
Sets the offset Z0
*1
*1.Specify 1 Ω when Waveguide is set for the media type.
Offset Loss Sets the offset loss
Arb. Impedance Sets an arbitrary impedance
Min. Frequency
Sets a start frequency.
*2
*2.Specify cutoff frequency when Waveguide is set for the media type.
Max. Frequency
Sets a stop frequency.
*2
Connector Type Selects a media type.