Chapter 12 401
Optimizing Measurements
Improving Phase Measurement Accuracy
12. Optimizing Measurements
Step 3. Press .
Step 4. Press
Marker Delay to set the electrical delay to the group delay value at the position of
the active marker (a value smoothed with the aperture of 20% regardless of the smoothing
setting).
NOTE An absolute value is used regardless of the reference marker mode’s on/off status.