Agilent Technologies ENA Series Guia do Utilizador Página 115

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 114
Chapter 4 109
Calibration
OPEN/SHORT Response Calibration (reflection test)
4. Calibration
Step 8. Press Open or Short to start the calibration measurement.
Step 9. If an isolation calibration must be performed using a LOAD standard, follow the procedure
below.
a. Connect a LOAD standard to the test port (connector to which the DUT is to be
connected) selected in Step 6.
Figure 4-6 Connecting the LOAD standard
b. Press
Load (Optional) to start the measurement on the LOAD standard.
Step 10. Press
Done to terminate the response calibration (and the LOAD isolation calibration)
process. Upon pressing this key, calibration coefficients will be calculated and saved. The
error correction function will also be automatically enabled.