494 Chapter 15
Measurement Examples
Measuring the SAW Bandpass Filter Using the Segment Sweep
3. Select the Segment Sweep as the Sweep Type
The segment sweep is selected as the sweep type.
4. Execute the Calibration
In this step, a 2-port ECal is executed on the two ports to be used.
Step 1. Connect the ECal module across test ports 1 and 2.
Figure 15-3 Connecting the ECal module
Step 2. Execute the 2-port ECal.
Setup Description Key Operation
Sweep type: Segment sweep
-
Sweep Type - Segment
Setup Description Key Operation
Executing a 2-port ECal between test ports 1
and 2
-
ECal - 2 Port ECal - 1-2