Chapter 4 111
Calibration
THRU Response Calibration (transmission test)
4. Calibration
NOTE An asterisk (*) in the upper right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute acquisition of the
calibration coefficient (pressing Done).
Step 7. Make a THRU connection between the test ports (between the connectors to which the
DUT will be connected) selected in Step 6.
Figure 4-9 Connecting at THRU response calibration
3-1 (S31) Selects test port 3 (input) and test port 1 (output); Corresponds to the
determination of S31
4-1 (S41) Selects test port 4 (input) and test port 1 (output); Corresponds to the
determination of S41
1-2 (S12) Selects test port 1 (input) and test port 2 (output); Corresponds to the
determination of S12
3-2 (S32) Selects test port 3 (input) and test port 2 (output); Corresponds to the
determination of S32
4-2 (S42) Selects test port 4 (input) and test port 2 (output); Corresponds to the
determination of S42
1-3 (S13) Selects test port 1 (input) and test port 3 (output); Corresponds to the
determination of S13
2-3 (S23) Selects test port 2 (input) and test port 3 (output); Corresponds to the
determination of S23
4-3 (S43) Selects test port 4 (input) and test port 3 (output); Corresponds to the
determination of S43
1-4 (S14) Selects test port 1 (input) and test port 4 (output); Corresponds to the
determination of S14
2-4 (S24) Selects test port 2 (input) and test port 4 (output); Corresponds to the
determination of S24
3-4 (S34) Selects test port 3 (input) and test port 4 (output); Corresponds to the
determination of S34
Softkey Function