668 Appendix D
Softkey Functions
Calibration Menu
(Continued)
Calibrate
Response (Thru)
Displays softkeys for selecting options for response calibrations using the THRU standard.
Select Ports Displays softkeys for selecting test ports.
2-1 (S21) Selects the transmission test (measurement of S
21
) for test port
12.
:SENS{1-16}:CORR:COLL:METH:THRU 2,1
3-1 (S31)
*1
Selects the transmission test (measurement of S
31
) for test port
13.
:SENS{1-16}:CORR:COLL:METH:THRU 3,1
4-1 (S41)
*2
Selects the transmission test (measurement of S
41
) for test port
14.
:SENS{1-16}:CORR:COLL:METH:THRU 4,1
1-2 (S12)
Selects the transmission test (measurement of S
12
) for test port
21.
:SENS{1-16}:CORR:COLL:METH:THRU 1,2
3-2 (S32)
*1
Selects the transmission test (measurement of S
32
) for test port
23.
:SENS{1-16}:CORR:COLL:METH:THRU 3,2
4-2 (S42)
*2
Selects the transmission test (measurement of S
42
) for test port
24.
:SENS{1-16}:CORR:COLL:METH:THRU 4,2
1-3 (S13)
*1
Selects the transmission test (measurement of S
13
) for test port
31.
:SENS{1-16}:CORR:COLL:METH:THRU 1,3
2-3 (S23)
*1
Selects the transmission test (measurement of S
23
) for test port
32.
:SENS{1-16}:CORR:COLL:METH:THRU 2,3
4-3 (S43)
*2
Selects the transmission test (measurement of S
43
) for test port
34.
:SENS{1-16}:CORR:COLL:METH:THRU 4,3
1-4 (S14)
*2
Selects the transmission test (measurement of S
14
) for test port
41.
:SENS{1-16}:CORR:COLL:METH:THRU 1,4
2-4 (S24)
*2
Selects the transmission test (measurement of S
24
) for test port
42.
:SENS{1-16}:CORR:COLL:METH:THRU 2,4
3-4 (S34)
*2
Selects the transmission test (measurement of S
34
) for test port
43.
:SENS{1-16}:CORR:COLL:METH:THRU 3,4
Cancel
Returns to the softkey display screen one level higher.
Thru Executes a THRU response calibration on the test ports selected in
the
Select Ports menu. This calibration is effective for eliminating
transmission tracking errors.
:SENS{1-16}:CORR:COLL:THRU
Isolation
(Optional)
Executes an isolation calibration on the test ports selected in the
Select Ports menu. This calibration is effective for eliminating
isolation errors.
:SENS{1-16}:CORR:COLL:ISOL
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return Returns to the softkey display screen one level higher.
Enhanced
Response
Displays softkeys for executing enhanced response calibrations.
Select Port Displays softkeys for selecting a test port.
2-1 (S21 S11) Selects the transmission test (measurement of S
21
) for test port
12.
:SENS{1-16}:CORR:COLL:METH:ERES 2,1
Key Operation Function SCPI Command