Agilent Technologies ENA Series Guia do Utilizador Página 140

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 139
134 Chapter 4
Calibration
Full 3-Port and Full 4-Port Calibration using 2-Port ECal
Figure 4-27 ECalAssistant (Connection) dialog box
Step 14. Re-connect the ECal module following the instructions given in each dialog box and
continue the calibration process.
Step 15. When all calibration data have been collected, a dialog box with the Complete! sign
appears as shown in Figure 4-28. Press the
Done button to finish the calibration. If you
wish to cancel the calibration, press the
Cancel button.
Figure 4-28 EcalAssistant (Finish) dialog box
Vista de página 139
1 2 ... 135 136 137 138 139 140 141 142 143 144 145 ... 774 775

Comentários a estes Manuais

Sem comentários