422 Chapter 12
Optimizing Measurements
Performing a Segment-by-Segment Sweep (segment sweep)
Step 5. Enter each item in the following table for each added segment (line) to create the segment
table.
To create the segment table, use the following keys.
Start Sets the start value of the sweep range
Stop Sets the stop value of the sweep range
Center Sets the center value of the sweep range
Span Sets the span value of the sweep range
Points Sets the number of points
IFBW Sets the IF bandwidth
Power Sets the power level; the power range is common to the settings for the
linear/log sweep ( -
Power Ranges)
Delay Sets the sweep delay time
Sweep Mode Sets the sweep mode; you need to select one of the following items:
STD STEPPED Stepped mode
STD SWEPT Swept mode
FAST STEPPED Fast stepped mode
FAST SWEPT Fast swept mode
Time Sets the sweep time; to specify the auto setting (AUTO), enter 0 as the sweep
time
Hardkey Function
If you select a cell and then press this key, you enter the mode that allows
you to edit the cell character by character. If you change a value and then
press this key, the value is entered in the cell.
Moves up and down in the cell selected in the segment table. In the
character-by-character edit mode, you can select an item or perform a
stepwise change of data.
Moves left and right in the cell selected in the segment table.
Softkey Function
Delete Deletes the line containing the selected cell.
Add Adds a new line above the line containing the selected cell.
Clear Segment Table -
OK
Resets the segment table. As a result, the initialized segment 1 remains.