Chapter 4 199
Calibration
Changing the Calibration Kit Definition
4. Calibration
85050C/D
NOTE The designation (m) in the name (label) of the standard indicates male analyzer connectors.
1. Short(m) 2. Open(m) 3. Broadband 4. Thru
Label Short Open Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0 90.48 m 0 0
C1 [×10
-27
F/Hz]
0 763.3 m 0 0
C2 [×10
-36
F/Hz
2
]
0 -63.82 m 0 0
C3 [×10
-45
F/Hz
3
]
0 6.434 m 0 0
L0 [×10
-12
H]
356.6 m 0 0 0
L1 [×10
-24
H/Hz]
-33.39 0 0 0
L2 [×10
-33
H/Hz
2
]
1.754 0 0 0
L3 [×10
-42
H/Hz
3
]
-33.60 m 0 0 0
Offset Delay [s] 0 0 0 0
Offset Z0 [Ω]50505050
Offset Loss [Ω/s] 0 0 0 0
Arb. Impedance [Ω]50507550
5. 2-18 Line 6. Lowband
Label Thru Lowband
STD Type Delay/Thru Load
C0 [×10
-15
F]
00
C1 [×10
-27
F/Hz]
00
C2 [×10
-36
F/Hz
2
]
00
C3 [×10
-45
F/Hz
3
]
00
L0 [×10
-12
H]
00
L1 [×10
-24
H/Hz]
00
L2 [×10
-33
H/Hz
2
]
00
L3 [×10
-42
H/Hz
3
]
00
Offset Delay [s] 23.190 p 0
Offset Z0 [Ω] 49.988 50
Offset Loss [Ω/s] 0.7 G 0
Arb. Impedance [Ω]50 50