Agilent Technologies ENA Series Guia do Utilizador Página 549

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 548
Chapter 16 543
Specifications and Supplemental Information
Corrected System Performance
16. Specifications and
Supplemental Information
Table 16-5 Corrected System Performance With 3.5 mm Device Connector Type, 85093C
Electronic Calibration Module
Network analyzer: E5070B/E5071B, Calibration module: 85093C (3.5 mm, 50 Ω) electronic calibration (ECal)
module, Calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation
from calibration temperature, Isolation calibration not omitted
Description
Specification (dB)
3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz
Directivity 52 51 47
Source Match 44 39 34
Load Match 47 44 40
Reflection Tracking ±0.030 ±0.050 ±0.070
Transmission Tracking ±0.039 ±0.069 ±0.117
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)