154 Chapter 4
Calibration
Unknown Thru Calibration
Procedure to select the standard type and define the standard factor
Step 1. Press .
Step 2. Press
Modify Calkit.
Step 3. Press
Define STDs.
Step 4. Press
{Defined Name}.
Step 5. Press
STD Type.
Step 6. Press
Unknown Thru.
Step 7. Press
Offset Delay, then enter approximate offset delay value of the unknown thru
standard.
Procedure to define the unknown thru
Step 1. Press .
Step 2. Press
Modify Calkit.
Step 3. Press
Specify CLSs.
Step 4. Press
Thru.
Step 5. Press
Set All.
Step 6. Press
{Defined Name}.
NOTE Different unknown thru can be used for different ports.
Procedure to perform the unknown thru
Step 1. Press .
Step 2. Press
Calibrate.
Step 3. Press
n-Port Cal
*1
.
Step 4. Press
Transmission.
Step 5. Press
{Defined Name} to execute mechanical unknown thru calibration.
Step 6. Complete Reflection, and Isolation (optional) for n-full port calibration.
*1.The 1-port calibration is not available.