Agilent Technologies ENA Series Guia do Utilizador Página 508

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 507
502 Chapter 15
Measurement Examples
Evaluating a Duplexer
6. Define the Setup for Display
The setup for display is defined.
7. Execute the Measurement
A trigger is applied to execute the measurement.
8. Define the Setup for the Segment Display and Scale
The setup for the scale is defined.
Setup Description Key Operation
Number of traces to be displayed: 5
-
Number of Traces - 5
Trace placement: Trisected
Allocate Traces -
Measurement Parameter
Trace 1: S
13
- S13
Trace 2: S
21
- - S21
Trace 3: S
23
- - S23
Trace 4: S
33
- - S33
Trace 5: S
11
- - S11
Setup Description Key Operation
Trigger mode: Single (or continuous)
- Single (or Continuous)
Setup Description Key Operation
Segment display: Order base
-
Segment Display - Order
Base
Reference Line Position
Trace 1: 10
- - Reference Position -
Trace 2: 10
- Reference Position -
Trace 3: 10
-
Reference Position -
Trace 4: 9
- Reference Position -
Trace 5: 9
- Reference Position -