2-12 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
Pulsed Sweep Measurements
Pulsed sweep measurement is performed as shown below. The source channel forces
pulsed sweep voltage or current, and the measurement channel performs one point
measurement at each sweep step.
Figure 2-7 Pulsed Sweep Measurements
1. The pulsed sweep source sets output by the PT, WM, and PWV or PWI
commands. Only one channel can be used for the pulsed sweep source.
2. The pulsed sweep source starts output by a trigger, such as the XE command.
3. After the hold time, the measurement channel starts measurement as shown in
Figure 2-7. Only one channel can be used for measurement. Measurement time
is set by the AIT 2 command.
4. After measurement, the pulsed sweep source forces the pulse base value, and
waits for the rest of the pulse period. Then the pulsed sweep source changes the
output value.
5. The B1500 repeats measurement and 4 for all sweep steps.
6. After the pulsed sweep measurement, the pulsed sweep source forces the pulse
base value, and keeps it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
Voltage or current
Stop value
Start value
Previous value
Time
: Measurement
Hold time
Pulse period
Trigger (e.g. XE)
PT/WM/PWV/PWI
Pulse base value
Pulse width
Pulse period Pulse period
Measurement time (set by AIT 2)