Agilent Technologies B1500A Manual de Serviço Página 104

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 598
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 103
2-18 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
To Stop Sweep
Output
An automatic abort function is available. Refer to “Automatic Abort Function” on
page 2-46.
Even if the automatic abort function is disabled, the B1500 automatically stops
measurement if power compliance is enabled for the sweep source and the power
compliance or an automatic abort condition is detected.
Quasi-Pulsed Spot Measurements
Quasi-pulsed spot measurement is performed as shown below. The measurement
channel performs one point measurement while the source channel forces a
quasi-pulse voltage. This measurement mode can minimize the output time of the
measurement voltage. So it is effective for the breakdown voltage measurement and
the reliability test.
Figure 2-11 Quasi-Pulsed Spot Measurements
1. The quasi-pulse source channel sets output by the BDT, BDM, and BDV
commands. Only one channel can be used for the quasi-pulse source.
2. The quasi-pulse source starts output by a trigger, such as the XE command.
3. After the hold time, the quasi-pulse source starts the voltage transition to the
stop value (settling detection time). Also, it performs voltage measurement
(settling detection) in the interval set by the BDM command. The voltage
transition and settling detection continue until the output voltage slew rate
becomes half of the rate when settling detection started. The slew rate depends
on the cabling and the characteristics of the device. You cannot define it directly.
In normal operation, the slew rate will be slower in the following conditions:
Voltage
Measurement time
X value
Start value
Previous value
Time
: Measurement
Hold time
Delay
time
Trigger (e.g. XE)
BDT/BDM/BDV
Settling detection time
The X value is the voltage when the settling detection is stopped.
Vista de página 103
1 2 ... 99 100 101 102 103 104 105 106 107 108 109 ... 597 598

Comentários a estes Manuais