4-74 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
DO
Remarks • Before executing DIAG? 1, connect a BNC cable between the Ext Trig In and
Out connectors.
• After executing DIAG? 3, confirm the status of LED. Then enter the AB
command. result returns 2.
If the LED does not blink, the B1500 must be repaired.
• Before executing DIAG? 4, disconnect any cable from the digital I/O port.
• Before executing DIAG? 6, open interlock circuit.
• Before executing DIAG? 7, close interlock circuit.
Example
Statements
OUTPUT @B1500;"DIAG? 1"
ENTER @B1500;A
DO
The DO command executes the B1500 internal memory programs (up to 8
programs) in the order specified.
Execution
Conditions
The specified programs have been stored by using the ST and END commands.
Syntax DO pnum[,pnum[,pnum[,pnum[,pnum[,pnum[,pnum[,pnum]]]]]]]
Parameters pnum : Internal memory program number. Numeric expression. 1 to 2000.
Example
Statements
OUTPUT @B1500;"DO 1,2,3,4,5,6,7,8"
OUTPUT @B1500;"DO 98,99"
DSMPLARM
Available for the multi channel pulsed spot measurement (MM27) and the multi
channel pulsed sweep measurement (MM28) using HVSMU, HCSMU, MCSMU,
DHCSMU, UHCU, HVMCU, or UHVU.
The DSMPLARM command sets the timing to start the signal monitor operation set
by the DSMPLSETUP command. The signal monitor is started when the specified
event is repeated the specified count.
Execution
Conditions
The DSMPLSETUP command must be executed before this command.
Syntax DSMPLARM chnum,type,count