2-40 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
CV (AC level) Sweep Measurements
CV (AC level) sweep measurement is performed as shown below. The CMU
(capacitance measurement unit) applies AC signal with DC bias, and performs one
point measurement at each step of AC signal level sweep. While the sweep
measurement, the AC signal frequency and the DC bias are constant.
Before performing the measurement, select the measurement parameters by using
the IMP command. And select the output data by using the LMN command.
Figure 2-25 CV (AC level) Sweep Measurements
1. The CMU sets the AC signal level sweep output by the WTACV, WMACV,
WACV, and FC commands.
2. The CMU starts DC bias output by the DCV command.
3. The CMU starts AC signal level sweep by a trigger, such as the XE command.
4. After the hold time, the CMU waits for the delay time.
5. After the delay time, the CMU starts measurement.
6. After measurement, the CMU waits for the rest of the step delay time if it is set,
and the CMU changes the AC signal level.
7. The B1500 repeats 5 and 6 for all sweep steps.
8. After the sweep measurement, the CMU keeps the signal output of the AC signal
level’s start or stop value as specified by the WMACV command.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the CMU output to 0 V for both AC and DC.
AC signal level
Stop value
Start value
Previous value
Time
: Measurement
Hold time
Step delay time
Trigger (e.g. XE)
WTACV/WMACV/WACV/FC/DCV
Delay time
Step delay time
Delay time