Agilent Technologies B1500A Manual de Serviço Página 431

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 598
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 430
Agilent B1500A/B1505A Programming Guide, Edition 11 4-115
Command Reference
*LRN?
For the SPGU, the query response is:
CNX chnum[,chnum][;CL chnum]<CR/LF^EOI>
or
CL chnum[,chnum][;CNX chnum]<CR/LF^EOI>
where chnum of CNX is the channel number for the channel whose
output switch is set to ON, and chnum of CL is the channel number for
the channel whose output switch is set to OFF.
30 : Returns the filter ON/OFF status:
FL0 [off ch[,off ch . . . [,off ch] . . . ];
FL1 [on ch[,on ch ...[,on ch] . . ] <CR/LF^EOI>
If all modules are Filter OFF, the query response is:
FL0<CR/LF^EOI>
If all modules are Filter ON, the query response is:
FL1<CR/LF^EOI>
31 : Returns the parameter values of the TM, AV, CM, FMT, and MM
commands:
TM trigger mode;AV number[,mode];CM auto calibration mode;
FMT output data format,output data mode
[;MM measurement mode[,chnum[,chnum...[,chnum]...]]]
<CR/LF^EOI>
32 : Returns the measurement ranging status:
RI chnum,Irange;RV chnum,Vrange
[;RI chnum,Irange;RV chnum,Vrange]....
[;RI chnum,Irange;RV chnum,Vrange]<CR/LF^EOI>
33 : Returns the staircase sweep measurement settings:
WM automatic sweep abort function,output after sweep
;
WT hold time,delay time[,step delay time[,S trig delay[,M trig delay]]]
[;WV chnum,mode,range,start,stop,nop[,Icomp [,pcomp]]] or
[;WI chnum,mode,range,start,stop,nop[,Vcomp[,pcomp]]]
[;WSV chnum,range,start,stop[,Icomp[,pcomp]]] or
[;WSI chnum,range,start,stop[,Vcomp[,pcomp]]]<CR/LF^EOI>
34 : Returns the pulsed source settings:
Vista de página 430
1 2 ... 426 427 428 429 430 431 432 433 434 435 436 ... 597 598

Comentários a estes Manuais

Sem comentários