3-32 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
Staircase Sweep with Pulsed Bias Measurements
Staircase Sweep with Pulsed Bias Measurements
To perform staircase sweep with pulsed bias measurements, use the following
commands.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Measurement time AIT 2,mode[,N]
Sets auto abort function [WM] abort[,post]
Sets voltage sweep source WV chnum,mode,range,start,stop,
step[,comp[,Pcomp]]
Sets current sweep source WI
Sets synchronous sweep
source
a
a. The WSV/WSI command must be entered after the WV/WI command.
[WSV] chnum,range,start,stop
[,comp[,Pcomp]]
[WSI]
Sets pulse timing parameters PT hold,width,period [,tdelay]
Forces pulse voltage PV chnum,range,base,pulse[,comp]
Forces pulse current PI chnum,range,base,pulse [,comp]
Forces constant voltage DV, TDV chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current DI, TDI
Sets voltage measurement
range
[RV] chnum,range
Sets current measurement
range
[RI] chnum,range
[RM] chnum,mode[,rate]
Sets measurement mode MM 5,chnum
Sets SMU operation mode [CMM] chnum,mode
Executes measurement XE