Agilent Technologies B1500A Manual de Serviço Página 239

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 598
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 238
Agilent B1500A/B1505A Programming Guide, Edition 11 3-61
Programming Examples
Quasi-static CV Measurements
Quasi-static CV Measurements
To make quasi-static CV measurements, use the following commands.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets QSCV operation mode [QSC] mode
Sets offset cancel ON/OFF
or performs capacitance
offset measurement
[QSZ] mode
Sets voltage sweep source QSV chnum,mode,vrange,start,stop
,cvoltage,step[,Icomp]
Sets time parameters QST cinteg,linteg,hold,delay1[,delay2]
Sets leak current data output
and compensation ON/OFF
[QSL] data,compen
Sets measurement range [QSR] range
Sets QSCV smart operation [QSO] mode[,chnum[,Vcomp]]
Sets automatic abort function [QSM] abort[,post]
Sets measurement mode MM 13[,chnum]
Forces constant voltage [DV, TDV] chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current [DI, TDI]
Executes measurement XE