4-28 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
Table 4-13 UHCU Pulsed Current Source Setup Parameters
1
Table 4-14 UHCU Pulsed Voltage Source Setup Parameters
2
Table 4-15 UHVU Voltage Source Setup Parameters
3
Output range
(actually used)
Setting
resolution in A
current, start,
stop, base, bias, or
pulse in A
a
a. 0 A is valid for the pulse base value.
Maximum
Vcomp value in V
500 A 1E-3 0 to ± 500 ± 63
2000 A
b
b. Only for the N1265A-015.
4E-3 0 to ± 1500
1. The following commands are used for setting the UHCU to the current pulse source.
PI, PWI, MCPNX, MCPWNX
Output range
(actually used)
Setting
resolution in V
voltage, start, stop,
base, bias, or pulse
in V
a
a. 0 A is valid for the pulse base value.
Maximum
Icomp value in A
100 V 200E-6 0 to ± 60
± 500 or ± 1500
b
b. Only for the N1265A-015.
2. The following commands are used for setting the UHCU to the voltage pulse source.
PV, PWV, MCPNX, MCPWNX
Output range
(actually used)
Setting
resolution in V
voltage, start,
stop, base,
bias, or pulse
in V
Maximum Icomp
value in A
Pulse DC
10 kV 10E-3
0 to ± 10000
a
a. Pulse base and peak values must be the same polarity.
± 20E-3
b
b. Only for the pulse output within maximum 1 ms pulse width.
±10E-3
3. The following commands are used for setting the UHVU to the voltage source.
DV, TDV, WV, WSV, WNX, PV, PWV, LSV, LSSV, BSV, BSSV, MV, MCPNX, MCPWNX