4-96 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERPFGR
Query Response state<CR/LF^EOI>
Connection status. 0 (open) or 1 (connect).
Example
Statements
OUTPUT @B1500;"ERPFGP?"
ENTER @B1500;gpathUHC
ERPFGR
This command sets the connection status for the series resistor on the Gate path of
Agilent N1265A ultra high current expander/fixture.
Execution
Conditions
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
The ERPFGP 0 command has been already executed.
Syntax ERPFGR state
Parameters state : Connection status of the series resistor on the Gate path. Integer
expression. 0, 10, 100, or 1000.
0: 0 Ω
10: 10 Ω
100: 100 Ω
1000: 1000 Ω
Example
Statements
OUTPUT @B1500;"ERPFGR 100"
ERPFGR?
This query command returns the connection status for the series resistor on the Gate
path of Agilent N1265A ultra high current expander/fixture.
Syntax ERPFDS?
Query Response state<CR/LF^EOI>
Connection status of the series resistor on the Gate path. 0, 10, 100, or 1000 (see
ERPFGR).
Example
Statements
OUTPUT @B1500;"ERPFGR?"
ENTER @B1500;gateR