1-64 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Programming Tips
To Use Programs for Agilent 4142B
Agilent B1500 supports most of the commands and the data output format supported
by the Agilent 4142B Modular DC Source/Monitor. To reuse the programs created
for the Agilent 4142B, confirm the following and modify the programs if necessary.
To remove all unsupported commands
Some commands are not supported owing to differences in the modules
supported by each instrument. See Table 1-17 that shows the commands not
supported by the B1500. Do not use these commands.
Perform the linear search or binary search measurement as a substitute for the
analog search measurement that needs the analog feedback unit (AFU).
Use a source/monitor unit (SMU) instead of the voltage source/voltage monitor
unit (VS/VMU). Note that the SMU cannot perform the differential voltage
measurements.
FL command
The initial setting of the FL command is different. It is ON for the Agilent
4142B, and OFF for the B1500.
Add the FL1 command to use the filter.
AV command
This command is used to set the A/D converter of the B1500.
To set the high resolution ADC installed in the B1500, use the AAD and AIT
commands.
*TST? command
Use 11 to specify the B1500 mainframe instead of 9 that indicates the 4142B
mainframe.
Table 1-17 Modules and Commands Unsupported
Plug-in Module Commands
Analog Feedback Unit ASM, AT, ASV, AIV, AVI
High Current Unit PDM, PDI, PDV
High Voltage Unit POL
Voltage Source/Voltage Monitor Unit VM