4-202 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
TM
irange : Current measurement range or ranging type. Integer expression. See
Table 4-3 on page 4-19.
vrange : Voltage measurement range or ranging type. Integer expression. See
Table 4-2 on page 4-17.
If you do not specify the irange and vrange parameters, the channel uses the
minimum range that covers the compliance value and the minimum range that
covers the output value.
Example
Statements
OUTPUT @B1500;"TIV 1"
ENTER @B1500 USING "#,3X,12D,X";Idata
ENTER @B1500 USING "#,3X,12D,X";Vdata
PRINT "I=";Idata*1000;"mA, V=";Vdata*1000;"mV"
TM
The TM command specifies how events are effective for the following actions:
• Releasing the B1500 from the paused status set by the PA or PAX command
• Starting the measurement except for high speed spot measurement (when the
B1500 is not in the paused status set by the PA, PAX, WS, or WSX command)
Syntax TM mode
Parameters mode : Event mode. Integer expression. See below.
To enable the trigger function set by the TGP port,terminal,polarity,1 command,
the mode value must be 3.
mode Events
1 XE command and GPIB GET (Group Execute Trigger,
TRIGGER command in HP BASIC). Initial setting.
2 XE command
3 XE command and external trigger
4 XE command and MM command (automatic trigger after
the MM command execution)