Agilent Technologies B1500A Manual de Serviço Página 163

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 598
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 162
Agilent B1500A/B1505A Programming Guide, Edition 11 2-77
Remote Mode Functions
Trigger Function
Trigger Output
A trigger output operation example is shown in Figure 2-48. When the measurement
or source output setup is completed, the output trigger is sent through the port
specified by the TGP command. See Table 2-8.
Figure 2-48 Trigger Output Example, Staircase Sweep Measurement, Negative Logic
Initial Settings The following functions are available in the initial settings:
Trigger port: Ext Trig Out
Trigger type: Measurement Completion (type 1)
Commands for the trigger output: OS
Output Trigger An edge trigger will be sent or a gate trigger level will be returned when an
operation is completed (see Figure 2-49).
Initially, the negative edge trigger of approximately 10 μs width is sent.
For the measurements using SMU pulse, an edge trigger will be sent or a gate
trigger level will be returned at 10 μs before starting the pulse output. Trigger width
is approximately 100 μs for both edge trigger and gate trigger.
: Measurement
TGP t,2,p,3
XE
TGP t,2,p,2
TGP t,2,p,1
TGP t,2,p,3
TGP t,2,p,2
TGP t,2,p,1
Gate trigger
Edge trigger
Source trigger delay
Hold time
Delay time
Source trigger delay
Step delay time
Step Measurement Completion
Step Output Setup Completion
Measurement Completion
Step Measurement Completion
Step Output Setup Completion
Measurement Completion
Vista de página 162
1 2 ... 158 159 160 161 162 163 164 165 166 167 168 ... 597 598

Comentários a estes Manuais

Sem comentários