Agilent B1500A/B1505A Programming Guide, Edition 11 4-205
Command Reference
TSQ
Parameters mode : Time stamp function mode. Integer expression.
When the function is enabled, the B1500 returns the time data with the
measurement data. The time data is the time from timer reset to the start
of measurement. Refer to “Data Output Format” on page 1-25.
Remarks To read the time data with the best resolution (100 μs), the timer must be cleared
every 100 sec or less for the FMT1, 2, or 5 data output format, or
every 1000 sec or less for the FMT 11, 12, 15, 21, 22, or 25 data output format.
Example
Statements
OUTPUT @B1500;"TSC 1"
TSQ
The TSQ command returns the time data from when the TSR command is sent until
this command is sent. The time data will be put in the data output buffer as same as
the measurement data.
This command is effective for all measurement modes, regardless of the TSC
setting.
This command is not effective for the 4 byte binary data output format (FMT3 and
FMT4).
Syntax TSQ
Example
Statements
OUTPUT @B1500;"TSQ"
ENTER @B1500 USING "#,3X,13D,X";Time
PRINT "Time=";Time;"s"
TSR
This command clears the timer count. This command is effective for all
measurement modes, regardless of the TSC setting. This command is not effective
for the 4 byte binary data output format (FMT3 and FMT4).
Syntax TSR [chnum]
mode Description
0 Disables the time stamp function. Initial setting.
1 Enables the time stamp function.