4-228 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
WSV
Execution
Conditions
Available for the staircase sweep (MM 2), pulsed sweep (MM 4), or staircase sweep
with pulsed bias (MM5) measurement.
This command must be entered after the WV or PWV command that clears the
WSV command setting. The WI and PWI command also clears the WSV setting.
Syntax WSV chnum,range,start,stop[,Icomp[,Pcomp]]
Parameters chnum : SMU synchronous sweep source channel number. Integer expression. 1
to 10 or 101 to 1001. See Table 4-1 on page 4-16.
range : Ranging type for synchronous sweep voltage output. Integer
expression. See Table 4-4 on page 4-21.
The B1500 usually uses the minimum range that covers both start and
stop values to force the staircase sweep voltage. However, if you set
Pcomp and if the following formulas are true, the B1500 changes the
output range dynamically (20 V range or above). Range changing may
cause 0 V output in a moment. For the limited auto ranging, the
instrument never uses the range less than the specified range.
Icomp > maximum current for the output range
Pcomp/output voltage > maximum current for the output range
start, stop : Start or stop voltage (in V). Numeric expression. See Table 4-7 on page
4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or Table 4-15
on page 4-28 for each measurement resource type. start and stop must
have the same polarity for log sweep. Sweep mode, linear or log, and
the number of sweep steps are set by the WV or PWV command. If the
output voltage is greater than the allowable voltage for the interlock
open condition, the interlock circuit must be shorted.
Icomp : Current compliance (in A). Numeric expression. See Table 4-7 on page
4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or Table 4-15
on page 4-28 for each measurement resource type. If you do not set
Icomp, the previous value is used.
Compliance polarity is automatically set to the same polarity as the
output value, regardless of the specified Icomp. If the output value is 0,
the compliance polarity is positive.
If you set Pcomp, the maximum Icomp value for the measurement
resource is allowed, regardless of the output range setting.