Agilent Technologies B1500A Manual de Serviço Página 435

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 598
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 434
Agilent B1500A/B1505A Programming Guide, Edition 11 4-119
Command Reference
*LRN?
71 : Returns the MFCMU data output mode:
LMN mode<CR/LF^EOI>
72 : Returns the MFCMU’s ADC setting:
ACT mode, number<CR/LF^EOI>
73 : Returns the MFCMU measurement range:
RC chnum,mode,range<CR/LF^EOI>
80 : Returns the operation mode of the SCUU connection status indicator:
SSL chnum,mode<CR/LF^EOI>
81 : Returns the SCUU connection path:
SSP chnum,mode<CR/LF^EOI>
90 : Returns the MFCMU adjustment mode setting:
ADJ chnum,mode<CR/LF^EOI>
100 : Returns the CV (DC bias) sweep measurement settings:
WMDCV abort function[,output after sweep];
WTDCV hold,delay[,step delay[,S trig delay[,M trig delay]]]
[;WDCV chnum,mode,start,stop,step]<CR/LF^EOI>
101 : Returns the pulsed spot C measurement or pulsed sweep CV
measurement settings:
PTDCV hold,width[,period[,trig delay]]
[;PDCV chnum,base,peak] or
[;PWDCV chnum,mode,base,start,stop,
step]<CR/LF^EOI>
102 : Returns the C-f sweep measurement settings:
WMFC abort post;
WTFC hold,delay[,step delay[,S trig delay[,M trig delay]]]
[;WFC chnum,mode,start,stop,step]<CR/LF^EOI>
103 : Returns the CV (AC level) sweep measurement settings:
WMACV abort post;
WTACV hold,delay[,step delay[,S trig delay[,M trig delay]]]
[;WACV chnum,mode,start,stop,step]<CR/LF^EOI>
104 : Returns the C-t sampling measurement settings:
MTDCV h_bias,interval,number,h_base
[;MDCV chnum,base,bias,post]<CR/LF^EOI>
Vista de página 434
1 2 ... 430 431 432 433 434 435 436 437 438 439 440 ... 597 598

Comentários a estes Manuais

Sem comentários