4-76 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
DSMPLSETUP
If there is no monitor data, a dummy data is returned.
Example
Statements
OUTPUT @B1500;"DSMPLFLUSH 5"
DSMPLSETUP
Available for HVSMU, HCSMU, MCSMU, DHCSMU, UHCU, HVMCU, and
UHVU.
The DSMPLSETUP command sets the signal monitor function for the specified
channel. This function is used to monitor the channel input/output during the
measurement specified by the MM command. By using the DSMPLARM and
DSMPLFLUSH commands, the channel measures its voltage and current, and
repeats it the specified count in the specified interval. Then the voltage and current
measurements are performed at the same time.
Syntax DSMPLSETUP chnum,count,interval[,delay]
Parameters chnum : SMU channel number. Integer expression. 1 to 10. See Table 4-1 on
page 4-16.
count : Number of sampling. Integer expression. 1 to 4000 for HVSMU.
1 to 2000 for HCSMU, MCSMU, DHCSMU, UHCU, HVMCU, and
UHVU.
interval : Sampling interval. Numeric expression. 6 μs to 12 μs in 2 μs resolution
for HVSMU. 2 μs to 12 μs in 2 μs resolution for HCSMU, MCSMU,
DHCSMU, UHCU, HVMCU, and UHVU.
delay : Delay time. Numeric expression. - maximum count value × interval to
+ maximum count value × interval, in 2 μs resolution. Default setting is
0.
Set 110 μs to start the signal monitor at the timing to start the transition
from the pulse base value to the pulse peak value.
Example
Statements
OUTPUT @B1500;"DSMPLSETUP 5,100,6E-6,1E-4"
DV
The DV command forces DC voltage from the specified SMU.
Execution
Conditions
The CN/CNX command has been executed for the specified channel.