Agilent B1500A/B1505A Programming Guide, Edition 11 1-31
Programming Basics
Data Output Format
A Status. One character.
Status for Source_data: See Table 1-7 on page 1-32.
Severity of a status is W<E.
Status for measurement data: See Table 1-10 on page 1-33.
For SMU, the severity of a status is as follows:
For the quasi-pulsed spot measurement: N<T<C<V<X<G or S
For other measurement: N<G<S<T<C<V<X
B Channel number of the measurement/source channel. One character. See Table 1-11
on page 1-34.
C Data type. One character. See Table 1-12 on page 1-35.
D Measurement data, output data, time data, or sampling index. Twelve or 13 digits
depends on FMT setting, which may be one of the following:
sn.nnnnnEsnn or sn.nnnnnnEsnn
snn.nnnnEsnn or snn.nnnnnEsnn
snnn.nnnEsnn or snnn.nnnnEsnn
where,
s: Sign, + or -.
n: Digit, 0 to 9.
E: Exponent symbol.
E Status. Three digits. Ignore status for the Time value.
Status for Source_data: See Table 1-7 on page 1-32.
Severity of a status is W<E.
Status for measurement data:
For SMU status, see Table 1-8 on page 1-32.
For CMU status, see Table 1-9 on page 1-32.
If multiple status conditions are found, the sum of the EEE values is returned.
For example, if an A/D converter overflow occurred, and an SMU was
oscillating during the measurements, the returned EEE value is 3 (=1+2).