2-6 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
NOTE The PT command sets the pulse timing parameters, such as pulse width and pulse
period. The PV command sets voltage pulse, and the PI command sets current pulse.
The base and peak values must have the same polarity for the current pulse.
Multi Channel Pulsed Spot Measurements
Multi channel pulsed spot measurement is performed as shown below. The
measurement channels perform one point measurement while a source channel is
forcing a pulse. Up to ten channels can be used for both pulse source and
measurement. Both voltage output mode and current output mode are available for
the source channels.
Figure 2-3 Multi Channel Pulsed Spot Measurements by MM27,SMU1,SMU2,SMU3
1. Hold time, pulse period, measurement delay time, and number of measurements
are set by the MCPT command.
2. The pulse source is set by the MCPNT and MCPNX commands with the source
identification number N (N=1 to 10).
3. The DC bias output is started by the DV/DI command.
4. Pulse output is simultaneously started by a trigger, such as the XE command.
For HR/MP/HPSMU, available delay time value is 0. Also, the pulse width
value must be the same. If a different value is entered, the longest value is set.
5. The measurement channels perform measurements in parallel as shown in
Figure 2-3. Measurement time is set by the AIT 2 command. This setting is
effective for all measurement channels.
Voltage or current
Peak value
Base value
Previous value
Time
: Measurement
Hold time
Pulse width2
Pulse period
Trigger (e.g. XE)
MCPT/MCPNT/MCPNX
Delay time2
Measurement delay time
Measurement time (set by AIT 2)
Peak value
Base value
Previous value
Pulse width1
MCPNT/MCPNX
Delay time1
DV
Setup value
Previous value
SMU1
SMU2, N=1
SMU3, N=2
Trigger
Hold time
Delay time2
Delay time1