Agilent B1500A/B1505A Programming Guide, Edition 11 4-121
Command Reference
LSM
If start < stop, step must be positive, and if start > stop, step must be
negative. Maximum number of search steps is 1001.
Vcomp: Voltage compliance value (in V). Numeric expression. See Table 4-6 on
page 4-23, Table 4-8 on page 4-25, or Table 4-11 on page 4-27 for each
measurement resource type. If you do not specify Vcomp, the previous
value is set.
Example
Statements
OUTPUT @B1500;"LSI 1,0,0,1E-6,1E-8,10"
LSM
The LSM command enables or disables the automatic abort function for the linear
search measurement (MM14). The automatic abort function stops the search
operation when one of the following conditions occurs:
• Compliance on the measurement channel
• Compliance on the non-measurement channel
• Overflow on the AD converter
• Oscillation on any channel
This command also sets the post search condition for the linear search sources. After
the search measurement is normally completed, the linear search sources force the
value specified by the post parameter.
If the search operation is stopped by the automatic abort function, the linear search
sources force the start value after search.
Syntax LSM abort[,post]
Parameters abort : Automatic abort function. Integer expression.
1: Disables the function. Initial setting.
2: Enables the function.
post : Source output value after the search operation is normally completed.
Integer expression.
1: Start value. Initial setting.
2: Stop value.
3: Output value when the search target value is obtained.
If this parameter is not set, the search source forces the start value.