4-148 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
PAD
OUTPUT @B1500;"CN";1
OUTPUT @B1500;"WAT";1,0,1E-3 !Source Wait Time =1ms
OUTPUT @B1500;"WAT";2,0,1E-3 !Meas Wait Time =1ms
OUTPUT @B1500;"DV";1,0,5,1E-2
OUTPUT @B1500;"PA";1E-3 !Wait Time =1ms
OUTPUT @B1500;"TI";1
ENTER @B1500 USING "#,3X,13D,X";Idata
Example
Statements
OUTPUT @B1500;"PA 10"
See Also “TM”
PAD
Enables or disables parallel measurements by the multiple channels (SMU). This
command is effective for the SMUs that use the high speed A/D converter (ADC)
and for the spot measurement (MM1), staircase sweep measurement (MM2),
sampling measurement (MM10), or CV (DC bias) sweep measurement (MM18).
The parallel measurements are performed at first by using the SMUs that use the
high speed ADC. And the other SMUs perform measurements in series as defined in
the MM command by using the high resolution ADC.
Syntax PAD mode
Parameters mode : 1 (enable) or 0 (disable, initial setting). Integer expression.
Example
Statements
OUTPUT @B1500;"PAD 1"
PAX
The PAX command pauses the command execution or internal memory program
execution, until the specified wait time elapses or until an event specified by the TM
command is received. The event set by the TM command only releases the paused
status. It does not start the measurement.
Execution
Conditions
The port parameter is meaningful only for the event (trigger input) set by the TM3
command. Set 1 (dummy) for the event set by the TM1, TM2, or TM4 command.
Syntax PAX port[,wait time]
Parameters port : External trigger input port number. Integer expression. -1, or 1 to 16.
-1: Ext Trig In terminal.