Agilent Technologies B1500A Manual de Serviço Página 384

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 598
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 383
4-68 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
CORRST
OUTPUT @B1500;"SPV 101,1,-0.5,0.5"
OUTPUT @B1500;"CORRSER? 101,1,1E-7,1E-8,10"
ENTER @B1500;A,B
In this example, the voltage measurement data is returned to the variable B, and the
load impedance calculation data is returned to the variable A.
See Also “SER”, “SER?”
NOTE Terminal voltage measurement and load impedance calculation
SPGU performs voltage measurement and impedance calculation by executing the
CORRSER? command. Followings are the recommended measurement conditions
for the voltage measurement.
Output voltage: > 1 V
Minimum load impedance: 40 Ω
Maximum load impedance: 500 Ω (1 V), 2000 Ω (2 V), 5000 Ω (10 V)
Figure 4-3 Voltage Monitor and CORRSER? Setup Parameters
CORRST
This command enables or disables the MFCMU open/short/load correction function.
Before setting a function to ON, perform the corresponding correction data
measurement by using the CORR? command.
delay interval
time
Ch1 output
Ch2 output
Start measurement
Meas[1]
Meas[2]
Meas[n]
delay + int erval x n
Measurement value = average(Meas[n] )
1 pulse width or 1 ALWG sequence
Pulse period >
n: count
Vista de página 383
1 2 ... 379 380 381 382 383 384 385 386 387 388 389 ... 597 598

Comentários a estes Manuais

Sem comentários