Agilent B1500A/B1505A Programming Guide, Edition 11 4-99
Command Reference
ERPFUHCTST?
Execution
Conditions
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
Syntax ERPFUHCMAX?
Query Response imax<CR/LF^EOI>
If the N1265A has the 1500 A option (N1265A-015), 1500 is returned. If not, 500 is
returned.
Example
Statements
OUTPUT @B1500;"ERPFUHCMAX?"
ENTER @B1500;maxUHC
ERPFUHCTST?
This query command performs the self-test of Agilent N1265A ultra high current
expander/fixture, and returns the execution results. After this command, read the
result soon.
Execution
Conditions
The ERPFUHCA command has been executed before this command.
Syntax ERPFUHCTST?
Query Response result<CR/LF^EOI>
0: Pass
1: Fail
Example
Statements
OUTPUT @B1500;"ERPFUHCTST?"
ENTER @B1500;result
ERR?
The ERR? query command returns error code. This command supports the error
code 0 to 999. Use the ERRX? command to support all errors.
Syntax ERR? [mode]
Parameters mode : Error code output mode. Integer expression. 0 (default setting) or 1.
0: Reads the contents of the error queue, and clears the queue. This
returns four error codes in order from their occurrence.