4-58 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
*CAL?
Execution
Conditions
No SMU may be in the high voltage state (forcing more than the allowable voltage
for the interlock open condition, or voltage compliance set to more than it).
Before starting the calibration, open the measurement terminals.
Syntax *CAL? [slot]
Parameters slot : Slot number where the module under self-calibration has been installed.
1 to 10. Or 0 or 11. Integer expression.
0: All modules and mainframe. Default setting.
11: Mainframe.
If slot specifies the slot that installs no module, this command causes an error.
Query Response results<CR/LF^EOI>
results returns the sum of the following values corresponding to the failures.
Example
Statements
OUTPUT @B1500;"*CAL?"
ENTER @B1500;A
Remarks If a SMU connected to SCUU fails the self-calibration, the SCUU cannot be
controlled. And the SSP and SSL commands are not available.
Failed modules are disabled, and can only be enabled by the RCV command.
results Description results Description
0 Passed. No failure detected. 32 Slot 6 module failed.
1 Slot 1 module failed. 64 Slot 7 module failed.
2 Slot 2 module failed. 128 Slot 8 module failed.
4 Slot 3 module failed. 256 Slot 9 module failed.
8 Slot 4 module failed. 512 Slot 10 module failed.
16 Slot 5 module failed. 1024 Mainframe failed.