742 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
33 Obsolete and Discontinued Commands
:MTESt:RUMode:SOFailure
(see page 790)
Command Syntax
:MTESt:RUMode:SOFailure <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RUMode:SOFailure command enables or disables the Stop On
Failure run until criteria. When a mask test is run and a mask violation is
detected, the mask test is stopped and the acquisition system is stopped.
Query Syntax
:MTESt:RUMode:SOFailure?
The :MTESt:RUMode:SOFailure? query returns the current state of the Stop
on Failure control.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 379
":MTESt:RUMode" on page 741
NOTE
The :MTESt:RUMode:SOFailure command is obsolete and is provided for backward
compatibility to previous oscilloscopes. Use the :MTESt:RMODe:FACTion:STOP command
(see page 400) instead.