:SEARch Commands 26
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 537
:SEARch:SERial:CAN:MODE
(see page 790)
Command Syntax
:SEARch:SERial:CAN:MODE <value>
<value> ::= {DATA | IDData | IDEither | IDRemote | ALLerrors
| OVERload | ERRor}
The :SEARch:SERial:CAN:MODE command selects the type of CAN
information to find in the Lister display:
DATA - searches for CAN Data frames matching the specified ID, Data,
and the DLC (Data length code).
IDData - searches for CAN frames matching the specified ID of a Data
frame.
IDEither - searches for the specified ID, regardless if it is a Remote
frame or a Data frame.
IDRemote - searches for CAN frames matching the specified ID of a
Remote frame.
ALLerrors - searches for CAN active error frames and unknown bus
conditions.
OVERload - searches for CAN overload frames.
ERRor - searches for CAN Error frame.
Query Syntax
:SEARch:SERial:CAN:MODE?
The :SEARch:SERial:CAN:MODE? query returns the currently selected
mode.
Return Format
<value><NL>
<value> ::= {DATA | IDD | IDE | IDR | ALL | OVER | ERR}
See Also Chapter 26, “:SEARch Commands,” starting on page 531
":SEARch:SERial:CAN:PATTern:DATA" on page 538
":SEARch:SERial:CAN:PATTern:ID" on page 540