Agilent Technologies 2000 X-Series Manual de Serviço Página 685

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 930
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 684
:WGEN Commands 31
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 685
:WGEN:OUTPut:LOAD
(see page 790)
Command Syntax
:WGEN:OUTPut:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :WGEN:OUTPut:LOAD command selects the expected output load
impedance.
The output impedance of the Gen Out BNC is fixed at 50 ohms. However,
the output load selection lets the waveform generator display the correct
amplitude and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the
displayed amplitude and offset levels will be incorrect.
Query Syntax
:WGEN:OUTPut:LOAD?
The :WGEN:OUTPut:LOAD? query returns the current expected output load
impedance.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also "Introduction to :WGEN Commands" on page 665
Vista de página 684
1 2 ... 680 681 682 683 684 685 686 687 688 689 690 ... 929 930

Comentários a estes Manuais

Sem comentários